Product Name |
Automatic Sample Processing System-GenMix Pro |
Handling Time |
≤ 20 minutes for 96 samples |
Compatible Sampling Tubes |
-Diameter: 13-19mm; Height: 55-115mm
-Compatible with 5mL, l0mL and 20mL screwcap sampling tubes; -5-in-l, 10-in-l mixed sampling tubes direct load with cap;
-Direct loading with cap for sampling tubes with swab;
-Preset with conventional standard sampling tubes, ready for use with customized sampling tubes specifications |
Compatible Deep-Well Plates |
-6*16T standard ImL deep-well plates
-1*96T standard ImL deep-well plates
-Compatible with 3mL deep-well plates (including 3 * 8T 3mL large-system deep-well plates of Tianlong)
-Preset with conventional standard deep-well plates, ready for use with customized deep-well plate specifications |
Pipetting |
-Dual independent pipetting modules
-Pipetting volume: 5 -1000 μL
-Liquid level sensor
-Pipetting volume detection |
Pipetting Accuracy |
5 μL~50 μL: Er: ≤3.5%
50 μL~200 μL: Er: ≤2.5%
≥ 200 μL: Er: ≤2.0% |
Pipetting
Repeatability/CV |
5 μL~50 μL: CV: ≤ 2.5%
50 μL~200 μL: CV: ≤ 1.5%
≥ 200 μL: CV: ≤ 1.0% |
Smart Information System |
-Sampling tube information scanning (ID code & QR code)
-Deep-well plate status identification (open system) /information scanning (Tianlong system)
-Automatic barcode scanning of sample racks
-"Sample - rack - plate” information scanning, "sample to result" in closed-loop management
-Lis system link |
Touch Screen |
12.1 inch LCD touch screen |
Dimensions |
1130mm(L) x780mm(W) x 920mm(H) |
Power Supply |
Voltage: 100 - 240 V; Frequency: 50/60HZ; Rated power: 60OVA |
Relative Humidity |
35%RH-70%RH, non-condensing |
Atmospheric Pressure Range |
56-106Kpa (Altitude ≤ 4000m) |